A Simple D.C. Method for the Determination of the Electrical Resistance of Glass
JSGT 1955 V39 T193-T204
A simple apparatus for the measurement of the DC conductivity of glasses at low temperatures is described. In order to make observations at low currents a pH meter is used, a range of measurement of specific resistances from 108 ohm cm to 5×1012 ohm cm being possible using the instrument in its original form. A modification of the pH meter permits this range to be extended up to a specific resistance of 1016 ohm cm. Typical results obtained with the apparatus are given.
H. E. Taylor